Sierra Series 1610Flo-Multimeter™Portable Air Velocity MeterMarch 1991REV. BPart Number: IM-615 Harris Court, Building L · Monterey, CA 93940(408) 373
1610 Flo-Multimeter Instruction Manual Page 7Operation of the Series 1610 is straightforward. All controls onthe front panel are clearly designate
1610 Flo-Multimeter Instruction Manual Page 84. Rotate the main switch to the temperature scale. The Flo-MultimeterTM is now measuring temperatur
1610 Flo-Multimeter Instruction Manual Page 9As discussed in Section 3, PRINCIPLE OF OPERATION, the Accu-FloTM probe measures the “standard” veloc
1610 Flo-Multimeter Instruction Manual Page 108CALIBRATION8.1Calibration at SierraInstruments, Inc.To ensure the continuing high accuracy of your
1610 Flo-Multimeter Instruction Manual Page 119MAINTENANCE9.1Breakage or Damage ofAccu-Flo™ Probe9.2Ni-Cad BatteryMaintenanceIf the sensors of the
1610 Flo-Multimeter Instruction Manual Page 1210TROUBLE-SHOOTINGPROBLEM POSSIBLE CAUSE ACTIONVelocity or pressure A. Probe not oriented A. Orient
1610 Flo-Multimeter Instruction Manual Page 13PROBLEM POSSIBLE CAUSE ACTIONBatteries won’t charge A. Battery charger A. To determine if battery(“B
1610 Flo-Multimeter Instruction Manual Page 14Important NoteThe Series 1610 Accu-FloTM probes are individually matched toeach meter. Do not inter
1610 Flo-Multimeter Instruction Manual Page 15Item # Description Part No.1 Bottom Enclosure 6102-0572 Sliding Door 6102-0584 Probe Entrance Plate
1610 Flo-Multimeter Instruction Manual Page 16Item # Description Part No.19 Pan Hd Phil Scr20 Zero Adjust21 Meter Movement22 Pan Hd Phil Scr23 Kep
1610 Flo-Multimeter Instruction Manual Page 19CUSTOMER CAUTIONRE: OXYGEN SERVICESierra Instruments, Inc., is not liable for any dam-age or persona
1610 Flo-Multimeter Instruction Manual Page 17APPENDIX BSchematic
1610 Flo-Multimeter Instruction Manual Page 181 Introduction ... 12 Descri
1610 Flo-Multimeter Instruction Manual Page 1This instruction manual applies to the following Sierra Instru-ments, Inc., Series 1610 Flo-Multimete
1610 Flo-Multimeter Instruction Manual Page 2Sierra’s Series 1610 Flo-MultimeterTM is a lightweight, portableand versatile instrument for measurin
1610 Flo-Multimeter Instruction Manual Page 3Sierra’s Accu-FloTM air velocity probe has two sensors, a velocitysensor and a temperature sensor (se
1610 Flo-Multimeter Instruction Manual Page 4Velocity Accuracy:± 2% of full scale over –20 to 60°C and 5 to 30 psia;± 4% of full scale over –55 to
1610 Flo-Multimeter Instruction Manual Page 5Figure 2Operating Timefor Batteries10000800060004000200004.5 5
1610 Flo-Multimeter Instruction Manual Page 6Inspect the air velocity sensor (the sensor with the small 0.032inch diameter). If it is visibly dir